The focus layer being used is controlled by the scan mode which is in use.
The
Reflective mode always places the focus at the surface of the glass plane.
The
Wide Transparency mode also places the focus at the surface of the glass plane and scans the entire width of the transparency unit. Please note that the transparency area is slightly smaller than the glass plane. The maximum resolution in this mode is limited and it can not offer infrared correction.
The normal
Transparency mode places the focus at the hight of the film holder. A film holder should always being used while this mode is active. This mode offers the maximum resolution and infrared correction.
You can further control the focus by adjusting the hight of the film holder.
The
V700/750 holders have feet which can be switched to two different settings. This is changed by either inserting the feet so the arrow points towards the normal (0) or the slightly higher (+) position.
The
V800/850 holders have sliders which allow to place the holder in a slightly lower or in three higher positions. The standard position is indicated by the arrow shaped notch.
Also make sure that you have selected a sufficiently high resolution.
The Resolution defines the amount of details in your scanned image.
Additionally you can use the
USM (UnSharpening Mask) to sharpen the images software sided.